Multiple MDT LNQ22 Transmissions Per Day
MDT LNQ22 devices generate dynamic reports that can affect how they appear on CareLink and Rhythm Science. This causes multiple reports to generate within R360 that may look the same but are slightly different.
When a non-Linq device transmits, the data becomes available in the CareLink dashboard for each transmission. When a LNQ-22 device transmits, the latest report on CareLink is updated each time creating a dynamic report and at the summary mark a summary report as well. Each will have a slightly different date and time because it's a real time updated EGM.
Our recommendation is read through the different reports due to the differences in each. If no new episodes exist, they can be noted accordingly and dismissed.